Page 100 - Exploring the Potential of Self-Monitoring Kidney Function After Transplantation - Céline van Lint
P. 100

 98
Chapter 4
SUPPLEMENTAL DATA
Table 1. Calibration performance of the creatinine Statsensor® device as determined by a clinical chemistry technician during the evaluation study. To that end, 5 calibrator levels were measured in duplicate on 6 different StatSensor® devices with two batches of strips. Averages, SD and overall CVs were calculated using basic statistics.
Creatinine
Level1
Level2
Lot
Calibrators
5511017241
5511017242
Range
μmol/L
71-150
133-248
141
207
Lot strips
4910348249
Mean μmol/L
SD μmol/L
5
11
n=12
VC%
3.2
5.2
131
200
Lot strips
   4911013249
Mean μmol/L
SD μmol/L
9
12
n=12
     VC%
6.5
 6.2
 Level3
Level4
Level5
5511017243
5511017244
5511017245
256-442
433-698
619-1061
391
558
816
26
20
20
6.8
3.6
2.5
382
564
807
21
46
27
5.6
 8.2
 3.3
                             